Type: Article
Publication Date: 2015-09-09
Citations: 69
DOI: https://doi.org/10.1103/physreva.92.031802
The temperature dependence of the asymmetry between Stokes and anti-Stokes Raman scattering can be exploited for self-calibrating, optically based thermometry. In the context of cavity optomechanics, we observe the cavity-enhanced scattering of light interacting with the standing-wave drumhead modes of a ${\mathrm{Si}}_{3}{\mathrm{N}}_{4}$ membrane mechanical resonator. The ratio of the amplitude of Stokes to anti-Stokes scattered light is used to measure temperatures of optically cooled mechanical modes, down to the level of a few vibrational quanta. We demonstrate that the Raman-ratio technique allows the measurement of the physical temperature of our device over a range extending from cryogenic temperatures to within an order of magnitude of room temperature.