Prefer a chat interface with context about you and your work?
In situ measurement of relative attenuation length of gadolinium-loaded liquid scintillator using source data at RENO experiment
Abstract We present in situ measurements of the relative attenuation length of the gadolinium-loaded liquid scintillator in the RENO (Reactor Experiment Neutrino Oscillation) detectors using radioactive source calibration data. We observed a steady decrease in the attenuation length of the Gd-LS in the RENO detectors by ∼50% in about four …