Type: Article
Publication Date: 2014-10-29
Citations: 6
DOI: https://doi.org/10.3762/bjnano.5.200
We present a theoretical framework for the dynamic calibration of the higher eigenmode parameters (stiffness and optical lever inverse responsivity) of a cantilever. The method is based on the tip–surface force reconstruction technique and does not require any prior knowledge of the eigenmode shape or the particular form of the tip–surface interaction. The calibration method proposed requires a single-point force measurement by using a multimodal drive and its accuracy is independent of the unknown physical amplitude of a higher eigenmode.
Action | Title | Year | Authors |
---|---|---|---|
+ PDF Chat | Determining surface properties with bimodal and multimodal AFM | 2014 |
Daniel Forchheimer Stanislav S. Borysov Daniel Platz David B. Haviland |